Validation and method comparison for a point-of-care lateral flow assay measuring equine whole blood insulin concentrations

2Citations
Citations of this article
11Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

The Wellness Ready Test (WRT) is a lateral flow, stall-side assay that measures equine insulin in whole blood and requires validation before recommending clinical use. We evaluated intra- and inter-assay precision and linearity and compared the WRT with a radioimmunoassay (RIA). Tested concentrations ranged from <139 to >695 pmol/L (<20 to >100 μIU/mL). For 20 replicates at each insulin level, intra-assay CVs of the WRT for insulin were 13.3%, 12.9%, and 15.3% at low (139–278 pmol/L; 20–40 μIU/mL), intermediate (278–417 pmol/L; 40–60 μIU/mL), and high (>417 (Formula presented.) >60 μIU/mL) concentrations, respectively. For 10 replicates at each level (3 assay lots), inter-assay CVs were 15.9%, 11.0%, and 11.7%, respectively. In the weighted linear regression of 5 measured insulin concentrations against expected concentrations, R2 = 0.98, slope = 1.02, and y-intercept = 14.4 pmol/L (2.08 μIU/mL). The Spearman correlation coefficient (rs) was 0.90 (95% CI: 0.85–0.94) between the WRT and RIA; the WRT = f(RIA) Passing–Bablok regression yielded the fit, y = 1.005x + 24.3 pmol/L (3.50 μIU/mL). The WRT result averaged 10.4% higher than the RIA result, with targeted bias of 25.9, 26.1, and 26.7 pmol/L (3.74, 3.76, and 3.84 μIU/mL) for cutoffs used to diagnose insulin dysregulation of 312, 347, and 451 pmol/L (45, 50, and 65 μIU/mL). Assay clinical sensitivities, specificities, and accuracies determined at the 3 selected clinical cutoffs and using the RIA as gold standard were 87–95%, 92–96%, and 91–95%, respectively (n = 99 samples). Observed total error was 28.4–30.4%. The WRT had acceptable precision, excellent linearity, and good association with the RIA.

Cite

CITATION STYLE

APA

Berryhill, E. H., Urbina, N. S., Marton, S., Vernau, W., & Alonso, F. H. (2023). Validation and method comparison for a point-of-care lateral flow assay measuring equine whole blood insulin concentrations. Journal of Veterinary Diagnostic Investigation, 35(2), 124–131. https://doi.org/10.1177/10406387221142288

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free