APRON LEAKAGE TEST USING X-RAY FLUOROSCOPY IN RSUD M. NATSIR SOLOK IN 2021

  • Nansih L
N/ACitations
Citations of this article
19Readers
Mendeley users who have this article in their library.

Abstract

Apron is a lead apron to protect the body from radiation hazards, to reduce exposure to fluoroscopy radiation, made of lead as a shield. Based on observations, the number of aprons was seven, purchased in 2011 and 2012. The apron was placed on a small table and the wrong apron was found. Seven of the aprons, one of which was on the first apron, was suspected to have been damaged due to the wrong way of storing the apron and the researcher noticed kinks in the suspected apron. This type of research is quantitative, in the Radiology Unit of RSUD. Mohammad Natsir Solok from January 2021. Population 7 apron, sample 7 apron, data processing by measuring using tools available in fluoroscopy, data analysis compared with Lambert and McKeon (2001) theory. The results showed that the leak in the upper one apron was 51.54 mm2, the lower part 38.21 mm2 in the lower two apron 51.43 mm2, the third apron in the middle right front area was 21.12 mm2, in the lower right front area 1.664 ,87 mm2, at the rear center area 1,435.57 mm2, the top five rear area apron 6.97 mm2, at the rear center area 26.98 mm2 and the top left front six area apron 46.07 mm2. Of the seven aprons studied, there were leaks that exceeded the tolerance limit, namely on the third apron. Keyword : Apron, Leakage, Fluoroscopy

Cite

CITATION STYLE

APA

Nansih, L. A. (2022). APRON LEAKAGE TEST USING X-RAY FLUOROSCOPY IN RSUD M. NATSIR SOLOK IN 2021. Jurnal Teras Kesehatan, 4(2), 47–53. https://doi.org/10.38215/jtkes.v4i2.74

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free