We present a novel automated method for assessment of image alignment, applied to non-rigid registration of brain Magnetic Resonance Imaging data (MRI) for image-guided neurosurgery. We propose a number of robust modifications to the Hausdorff distance (HD) metric, and apply it to the edges recovered from the brain MRI to evaluate the accuracy of image alignment. The evaluation results on synthetic images, simulated tumor growth MRI and real neurosurgery data with expert- identified anatomical landmarks, confirm that the accuracy of alignment error estimation is improved compared to the conventional HD. The proposed approach can be used to increase confidence in the registration results, assist in registration parameter selection, and provide local estimates and visual assessment of the registration error. © Springer-Verlag Berlin Heidelberg 2008.

Cite

CITATION STYLE

APA

Fedorov, A., Billet, E., Prastawa, M., Gerig, G., Radmanesh, A., Warfield, S. K., … Chrisochoides, N. (2008). Evaluation of brain MRI alignment with the robust hausdorff distance measures. In Lecture Notes in Computer Science (including subseries Lecture Notes in Artificial Intelligence and Lecture Notes in Bioinformatics) (Vol. 5358 LNCS, pp. 594–603). https://doi.org/10.1007/978-3-540-89639-5_57