Stepwise formal modelling and reasoning of insulin infusion pump requirements

8Citations
Citations of this article
3Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

An insulin infusion pump (IIP) is a critical software-intensive medical device that infuses insulin satisfying patient needs under safety and timing constraints that are appropriate for the treatment of diabetes. This device is used by millions of people around the world. The USA Food and Drug Administration (FDA) has reported several recalls in which IIP failures were responsible for a large number of serious illnesses and deaths. The failures responsible for this harm to people who are dependent on external insulin were caused by the introduction of hardware or software design errors during the system development process. This paper presents an incremental proof-based development of an IIP. We use the Event-B modelling language to formalize the given system requirements. Further, the Rodin proof tools are used to verify the correctness of functional behaviour, internal consistency checking with respect to safety properties, invariants and events.

Cite

CITATION STYLE

APA

Singh, N. K., Wang, H., Lawford, M., Maibaum, T. S. E., & Wassyng, A. (2015). Stepwise formal modelling and reasoning of insulin infusion pump requirements. In Lecture Notes in Computer Science (including subseries Lecture Notes in Artificial Intelligence and Lecture Notes in Bioinformatics) (Vol. 9185, pp. 387–398). Springer Verlag. https://doi.org/10.1007/978-3-319-21070-4_39

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free