Antibacterial Envelope Is Associated With Low Infection Rates After Implantable Cardioverter-Defibrillator and Cardiac Resynchronization Therapy Device Replacement: Results of the Citadel and Centurion Studies

51Citations
Citations of this article
69Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Objectives This study sought to determine whether the nonabsorbable TYRX Antibacterial Envelope (TYRX) reduces major cardiovascular implantable electronic device (CIED) infections 12 months after implant. Background TYRX is a monofilament polypropylene mesh impregnated with minocycline and rifampin specifically designed to hold a CIED in place and elute antimicrobials over time. There are limited data on its ability to reduce CIED infections. Methods We prospectively enrolled patients who underwent generator replacement with an implantable cardioverter-defibrillator (ICD) or cardiac resynchronization therapy device (CRT), treated with TYRX. The primary endpoints were major CIED infection and CIED mechanical complications. Given the differences in infection rates among ICD and CRT patients, 3 different control populations were used: a published benchmark rate for ICD patients, and both site-matched and comorbidity-matched controls groups for CRT patients. Results Overall, a major CIED infection occurred in 5 of 1,129 patients treated with TYRX (0.4%; 95% confidence interval: 0.0% to 0.9%), significantly lower than the 12-month benchmark rate of 2.2% (p = 0.0023). Among the TYRX-treated CRT cohort, the major CIED infection rate was 0.7% compared with an infection rate of 1.0% and 1.3% (p = 0.38 and p = 0.02) in site-matched and comorbidity-matched control groups, respectively. Among the ICD group, the 12-month infection rate was 0.2% compared with the published benchmark of 2.2% (p = 0.0052). The most common CIED mechanical complication in study patients was pocket hematoma, which occurred in 18 of the 1,129 patients (1.6%; 95% confidence interval: 0.8 to 2.5), which is comparable with a published rate of 1.6%. Conclusions Use of TYRX was associated with a lower major CIED infection rate. (TYRX™ Envelope for Prevention of Infection Following Replacement With a CRT or ICD; [Centurion]; NCT01043861/NCT01043705)

Cite

CITATION STYLE

APA

Henrikson, C. A., Sohail, M. R., Acosta, H., Johnson, E. E., Rosenthal, L., Pachulski, R., … Simons, G. R. (2017). Antibacterial Envelope Is Associated With Low Infection Rates After Implantable Cardioverter-Defibrillator and Cardiac Resynchronization Therapy Device Replacement: Results of the Citadel and Centurion Studies. JACC: Clinical Electrophysiology, 3(10), 1158–1167. https://doi.org/10.1016/j.jacep.2017.02.016

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free