Biofunctionalization process of a-SiC:H surfaces applied to an interdigitated microelectrode array to detect enterotoxigenic Escherichia coli

2Citations
Citations of this article
3Readers
Mendeley users who have this article in their library.
Get full text

Abstract

A biofunctionalization process of a-SiC:H surfaces has been applied to an interdigitated microelectrode array (IMA) whose microelectrodes are covered by this material. The biofunctionalization process has been monitored stage by stage using Fourier transform infrared spectroscopy (FTIR), while its effects on the electrical behavior of the IMA were recorded in electrical impedance spectra. The process involves hydroxylation, silanization, generation of aldehyde groups, binding via protein A, immobilization of anti-Escherichia coli polyclonal antibodies, entrapping and detection of enterotoxigenic Escherichia coli (ETEC) in Luria-Bertani medium. The FTIR spectra confirm the success of the process. Regarding the performance of the IMA, although the detection of ETEC is successful and its percentage change in impedance reaches a value of 133.37% to 107 CFU/mL, some considerations may be taken into account to improve the sensitivity of the IMA by mean of the optimization of both the IMA design and the biofunctionalization process.

Cite

CITATION STYLE

APA

Herrera-Celis, J., Reyes-Betanzo, C., Orduña-Díaz, A., Pérez-Coyotl, A., Hernández-Flores, A., Morales-Chávez, J., & Culebro-Gomez, A. (2017). Biofunctionalization process of a-SiC:H surfaces applied to an interdigitated microelectrode array to detect enterotoxigenic Escherichia coli. In IFMBE Proceedings (Vol. 60, pp. 34–37). Springer Verlag. https://doi.org/10.1007/978-981-10-4086-3_9

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free