This work presents the modelling of impulsional pH variations in microvolume related to water-based electrolysis and hydrogen peroxide electrochemical oxidation using an Electrochemical Field Effect Transistor (ElecFET) microdevice. This ElecFET device consists of a pH-Chemical FET (pH-ChemFET) with an integrated microelectrode around the dielectric gate area in order to trigger electrochemical reactions. Combining oxidation/reduction reactions on the microelectrode, water self-ionization and diffusion properties of associated chemical species, the model shows that the sensor response depends on the main influential parameters such as: (i) polarization parameters on the microelectrode, i.e., voltage (Vp) and time (tp); (ii) distance between the gate sensitive area and the microelectrode (d); and (iii) hydrogen peroxide concentration ([H2O2]). The model developed can predict the ElecFET response behaviour and creates new opportunities for H2O2-based enzymatic detection of biomolecules. © 2014 by the authors; licensee MDPI, Basel, Switzerland.
CITATION STYLE
Diallo, A. K., Djeghlaf, L., Launay, J., & Temple-Boyer, P. (2014). Modelling of impulsional pH variations using ChemFET-based microdevices: Application to hydrogen peroxide detection. Sensors (Switzerland), 14(2), 3267–3283. https://doi.org/10.3390/s140203267