Improving the sensitivity and specificity of a bioanalytical assay for the measurement of certolizumab pegol

4Citations
Citations of this article
11Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Aim: In order to evaluate placental transfer of certolizumab pegol (CZP), a more sensitive and selective bioanalytical assay was required to accurately measure low CZP concentrations in infant and umbilical cord blood. Results & methodology: A new electrochemiluminescence immunoassay was developed to measure CZP levels in human plasma. Validation experiments demonstrated improved selectivity (no matrix interference observed) and a detection range of 0.032-5.0 μg/ml. Accuracy and precision met acceptance criteria (mean total error ?20.8%). Dilution linearity and sample stability were acceptable and sufficient to support the method. Conclusion: The electrochemiluminescence immunoassay was validated for measuring low CZP concentrations in human plasma. The method demonstrated a more than tenfold increase in sensitivity compared with previous assays, and improved selectivity for intact CZP.

Cite

CITATION STYLE

APA

Smeraglia, J., Silva, J. P., & Jones, K. (2017, August 1). Improving the sensitivity and specificity of a bioanalytical assay for the measurement of certolizumab pegol. Bioanalysis. Future Medicine Ltd. https://doi.org/10.4155/bio-2017-0124