We present the design and experimental results of a new MEMS-based endoscopic optical coherence tomography (OCT) probe. The uniqueness of this miniature OCT imaging probe is a two-axis MEMS mirror with through-silicon vias (TSVs) for interconnecting. The TSV interconnection enables ultracompact probe design, successfully reducing the probe size to only 2.6 mm in diameter. The MEMS mirror is actuated by an electrothermal actuator that is capable of scanning 16 at only 3.6 V DC. Two-dimensional and three-dimensional OCT images of microspheres embedded in PDMS and acute rat brain tissue have been obtained with this miniature probe in a time-domain OCT system. © 2011 Society of Photo-Optical Instrumentation Engineers (SPIE).
CITATION STYLE
Liu, L., Wu, L., Sun, J., Lin, E., & Xie, H. (2011). Miniature endoscopic optical coherence tomography probe employing a two-axis microelectromechanical scanning mirror with through-silicon vias. Journal of Biomedical Optics, 16(2), 026006. https://doi.org/10.1117/1.3533323