Soil contamination may occur in the upcoming decommissioning activities of the TRIGA2000 Reactor. Measurement of contaminant radioactivity, which can be performed using single-channel spectroscopy, is required in soil decontamination processes. This research develops a timer/counter system for single-channel spectroscopy using a microcontroller. The performance of the ATMega328P microcontroller Timer/Counter on Arduino has been tested for single-channel spectroscopy. Microcontroller's Timer/Counter 1 is used as a counter while Timer/Counter 2 is used as a timer. Tests include the linearity test, comparative test, and chi-square test. The test results show that the ATMega328P microcontroller Timer/Counter works well and can be used as the end of a single-channel spectroscopic system.
CITATION STYLE
Sulaksono, S. T., Sukmabuana, P., & Nagara, N. (2022). MICROCONTROLLER ATMEGA328P TIMER/COUNTER FOR SINGLE CHANNEL GAMMA SPECTROSCOPY. JURNAL TEKNOLOGI REAKTOR NUKLIR TRI DASA MEGA, 24(3), 125. https://doi.org/10.17146/tdm.2022.24.3.6699