Implantable neuromodulation devices that interface with the peripheral nervous system are a promising approach to restore functions lost to nerve damage. Existing nerve stimulation electrodes require direct contact with the target nerve and are associated with mechanical nerve damage and fibrous tissue encapsulation. Endovascularly delivered electrode arrays may provide a less invasive solution. Using a hybrid tissue conductor-neuron model and computational simulations, this study demonstrates the feasibility of delivering electrical stimulation of a peripheral nerve from a blood vessel in the vicinity of the target and predicts that the stimulation intensity required strongly depends on nerve-vessel distance and relative orientation, which are important factors to consider when screening candidate blood vessels for electrode implantation.
CITATION STYLE
Liu, J. Y., Grayden, D. B., Keast, J. R., & John, S. E. (2021). Computational Modeling of an Endovascular Peripheral Nerve Interface. In Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBS (pp. 5966–5969). Institute of Electrical and Electronics Engineers Inc. https://doi.org/10.1109/EMBC46164.2021.9630085
Mendeley helps you to discover research relevant for your work.