Diseño de medidor de conductividad electrolítica, para operar en el rango de conductividad eléctrica del tejido humano

0Citations
Citations of this article
5Readers
Mendeley users who have this article in their library.
Get full text

Abstract

A Van der Pauw electrolytic conductivimeter was development to calibrate an electrical resistivity probe. This conductivimeter is used in studies of early cancer screening because such studies require a measurement probe, i.e., tetra-polar probe, which needs to be calibrated. The standard calibration procedure uses electrolytic solutions of known electrical conductivities. The development Van der Pauw electrolytic conductivimeter should be used as primary standard in the probe calibration. A maximum measurements uncertainty of 1.64% was achieved. © Springer-Verlag Berlin Heidelberg 2007.

Cite

CITATION STYLE

APA

Jaimes, S. A., Valencia, J. P., & Miranda, D. A. (2008). Diseño de medidor de conductividad electrolítica, para operar en el rango de conductividad eléctrica del tejido humano. In IFMBE Proceedings (Vol. 18, pp. 456–459). Springer Verlag. https://doi.org/10.1007/978-3-540-74471-9_106

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free