Routine constancy checks of electron energy are often time consuming because of the necessity to measure a dose at two depths. A technique is described that uses a double-wedge shaped phantom positioned on a Profiler diode array for measuring an electron energy constancy metric similar to R(50). The double-wedge electron profiles are invariant to phantom alignment in the wedge direction, unlike single wedge techniques, and the sensitivity of the technique is similar to water-based depth-dose measurements over an energy range of 6 to 20 MeV. Reproducibility results ranging from 0.01 to 0.03 cm were achieved for measurements taken over the course of 1.5 yrs. The technique is efficient in that only one phantom setup is required for all electron energies. (c) 2003 American College of Medical Physics.
CITATION STYLE
Wells, D. M., Picco, P. J., & Ansbacher, W. (2003). Electron energy constancy verification using a double-wedge phantom. Journal of Applied Clinical Medical Physics / American College of Medical Physics, 4(3), 204–208. https://doi.org/10.1120/jacmp.v4i3.2516