A sensitive and accurate isocratic high-performance liquid chromagraphic method is developed for the determination of both metaisomer impurity and assay of valdecoxib drug substance. This method uses a Phenomenex Luna C18 (2) column, a mobile phase of 60:30:10 (v/v) 20mM NaH2PO4, methanol, and tetrahydrofuran, respectively, with UV detection. This method is validated and its stability-indicating capability is established by performing stress studies under acidic, basic, oxidation, light, humidity, and thermal conditions. Valdecoxib is well separated from its metaisomer impurity with a resolution of more than 2.0. The limit of detection of 0.007% is obtained for the metaisomer impurity, and the relative response factor is also determined. Repeatability is good, with a relative standard deviation of not more than 0.2% and 0.8% for the assay and impurity methods, respectively. A system suitability test is developed with acceptance criteria and the requirements are met throughout the method validation. The method is validated as robust to variations in chromatographic conditions.
CITATION STYLE
Karthikeyan, K., Saravanan, R., Rajeswari, R., & Pillai, K. C. (2009). Validation of single isocratic HPLC method for the assay of valdecoxib and determination of metaisomer impurity. Journal of Chromatographic Science, 47(4), 309–314. https://doi.org/10.1093/chromsci/47.4.309
Mendeley helps you to discover research relevant for your work.