A high resolution dynamic heart model based on averaged MRI data

43Citations
Citations of this article
27Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

We are in the process of constructing a high resolution, high signal to noise ratio (SNR) dynamic MRI dataset for the human heart using methodology similar to that employed to construct a low-noise standard brain at the Montreal Neurological Institute. Several high resolution, low SNR magnetic resonance images of 20 phases over the cardiac cycle were acquired from a single subject. Images from identical phases and temporally adjacent phases were registered, and the image intensities were averaged together to generate a high resolution, high SNR dynamic magnetic resonance image volume of the human heart. Although this work is still preliminary, and the results still demonstrate residual artifacts due to motion an sub-optimal alignment of interleaved image slices, our model has a SNR that is improved by a factor of 2.7 over a single volume, spatial resolution of 1.5 mm3, and a temporal resolution of 60 ms. © Springer-Verlag Berlin Heidelberg 2003.

Cite

CITATION STYLE

APA

Moore, J., Drangova, M., Wierzbicki, M., Barron, J., & Peters, T. (2003). A high resolution dynamic heart model based on averaged MRI data. Lecture Notes in Computer Science (Including Subseries Lecture Notes in Artificial Intelligence and Lecture Notes in Bioinformatics), 2878, 549–555. https://doi.org/10.1007/978-3-540-39899-8_68