Preconcentration and determination of heavy metals in water, sediment and biological samples

23Citations
Citations of this article
65Readers
Mendeley users who have this article in their library.

Abstract

In this study, a simple, sensitive and accurate column preconcentration method was developed for the determination of Cd, Cu and Pb ions in river water, urine and sediment samples by flame atomic absorption spectrometry. The procedure is based on the retention of the analytes on a mixed cellulose ester membrane (MCEM) column from buffered sample solutions and then their elution from the column with nitric acid. Several parameters, such as pH of the sample solution, volume of the sample and eluent and flow rates of the sample were evaluated. The effects of diverse ions on the preconcentration were also investigated. The recoveries were >95 %. The developed method was applied to the determination of trace metal ions in river water, urine and sediment samples, with satisfactory results. The 3δ detection limits for Cu, Pb and Cd were found to be 2, 3 and 0.2 μg dm-3, respectively. The presented procedure was successfully applied for determination of the copper, lead and cadmium contents in real samples, i.e., river water and biological samples.

Cite

CITATION STYLE

APA

Shirkhanloo, H., Mousavi, H. Z., & Rouhollahi, A. (2011). Preconcentration and determination of heavy metals in water, sediment and biological samples. Journal of the Serbian Chemical Society, 76(11), 1583–1595. https://doi.org/10.2298/JSC101024139S

Register to see more suggestions