Evaluation of a bracketing calibration-based isotope dilution liquid chromatography–tandem mass spectrometry candidate reference measurement procedure for 17α-hydroxyprogesterone in human plasma

8Citations
Citations of this article
5Readers
Mendeley users who have this article in their library.
Get full text

Abstract

A candidate reference measurement procedure (RMP) based on isotope dilution coupled with liquid chromatography–tandem mass spectrometry (ID–LC–MS/MS) was developed and validated for the quantification of 17α-hydroxyprogesterone (17-OHP) in human plasma. 17-OHP spiked with a deuterium-labeled internal standard was extracted from plasma by liquid–liquid extraction with 1 mL n-hexane/ethyl acetate (3:2, v/v). Reversed-phase chromatography and positive electrospray ionization were used in the ID–LC–MS/MS. Gradient elution coupled with use of a C18-packed ultrahigh-performance liquid chromatography column allowed complete baseline resolution of 17-OHP from its structural analogue desoxycorticosterone in 6 min. To determine the 17-OHP level in human plasma, a bracketing calibration method was used to give higher accuracy and precision. The limit of detection and the lower limit of the measuring interval for the candidate RMP were 2.1 pg/mL (6.4 pmol/L) and 4.6 pg/mL (13.9 pmol/L), respectively. Extraction recovery was determined to be (96.08 ± 3.03)% (n = 3). Imprecision (intra-assay and interassay) was 4.03% or less at 0.83, 15.19, 64.22, and 313.46 ng/mL (2.51, 45.97, 194.34, and 948.56 nmol/L, respectively). Recoveries ranged from 98.05% to 102.24%. When comparing our RMP results with those obtained with an established RMP via International Federation of Clinical Chemistry and Laboratory Medicine external quality assessment scheme for reference laboratories in laboratory medicine (RELA) samples, we found that the biases ranged from -1.99% to 3.08% against the targets. No interference was observed, and the linear response ranged from 0.47 to 958.63 ng/mL (1.42 to 2900.90 nmol/L). Moreover, the candidate RMP was used to measure the concentration of 17-OHP in human plasma and was compared with an immunoassay using 40 plasma samples. The performance of the method meets the needs of an RMP (total coefficient of variation of 5% or less and bias of 3.08% or less). This method can be used for reference material value assignment of 17-OHP in human plasma matrix. It could also serve as an accurate reference baseline for routine methods to increase the accuracy and precision of certain clinical laboratory measurements. [Figure not available: see fulltext.]

Cite

CITATION STYLE

APA

Zhang, Q., Zhang, L., Lin, H., Cai, Z., Yan, J., Wang, Q., … Huang, X. (2019). Evaluation of a bracketing calibration-based isotope dilution liquid chromatography–tandem mass spectrometry candidate reference measurement procedure for 17α-hydroxyprogesterone in human plasma. Analytical and Bioanalytical Chemistry, 411(27), 7095–7104. https://doi.org/10.1007/s00216-019-02086-5

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free