Inter-operator and inter-device reproducibility of shear wave elastography in healthy muscle tissues

4Citations
Citations of this article
11Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Purpose: The study aimed to assess whether the more limiting factor in reproducibility of shear wave elastography (SWE) would be the operator dependency or the incompatibility of different ultrasound (US) devices. The interrater agreement with less experienced operators was studied. Methods: A total of 24 healthy volunteers participated in the study (18 females, 6 males; range of age 27–55 years). SWE of biceps brachii (BB) and tibialis anterior (TA) muscles was performed on both sides from all participants in both longitudinal and transverse orientation of the transducer in respect to muscle fibers. Two operators repeated the SWE with two different US devices from different manufacturers (scanners 1 and 2). Results: Intraclass correlation coefficient between the two operators was 0.91 (CI 0.88–0.93) for scanner 1 and 0.81 (CI 0.74–0.86) for scanner 2, respectively. Instead, there were significant differences in the SWE measurements between the two scanners, emphasizing in transverse orientation of the transducer. In the transverse transducer orientation, the mean shear wave velocity (SWV) in TA was 1.45 m/s (standard deviation [SD] ± 0.35 m/s) with scanner 1 and 2.35 m/s (SD ± 0.83 m/s) with scanner 2 (p < 0.001). In BB, the mean transverse SWV was 1.49 m/s (SD ± 0.35 m/s) with scanner 1 and 2.29 m/s (SD ± 0.63 m/s) with scanner 2 (p < 0.001). In longitudinal transducer orientation, the mean SWV in TA was 3.00 m/s (SD ± 0.73 m/s) with scanner 1 and 3.26 m/s (SD ± 0.42 m/s) with scanner 2 (p = 0.050). In BB, the mean longitudinal SWV was 3.60 m/s (SD ± 0.77 m/s) with scanner 1 and 3.96 m/s (SD ± 0.62 m/s) with scanner 2 (p = 0.019). The presented mean values were obtained by operator 1, there were no significant differences in the SWE measurements performed by the two operators. Conclusion: The results implicate that the reproducibility of the SWE measurements depends rather on the used US device than on the operator. It is recommendable that clinics collect reference values with their own US device and consider threshold values presented in previous studies only directional.

Cite

CITATION STYLE

APA

Vuorenmaa, A. S., Siitama, E. M. K., & Mäkelä, K. S. (2022). Inter-operator and inter-device reproducibility of shear wave elastography in healthy muscle tissues. Journal of Applied Clinical Medical Physics, 23(9). https://doi.org/10.1002/acm2.13717

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free