Grip strength (GS) is widely used in various fields such as sports, rehabilitation, and geriatrics to assess muscle strength, and to diagnose sarcopenia and frailty in older adults. There is a potential for measurement differences among different dynamometers available, and studies comparing GS measurements by variable tools have conflicting results. The two most frequently used dynamometers are the Jamar hydraulic (Jamar) and spring-type hand grip dynamometers, and Jamar has not been compared to Jamar PLUS+ Digital (Jamar+) dynamometer in older adults. So, we aimed to assess GS measurements with the Jamar as the reference standard against Jamar+, and spring-type Takei T.KK. 5401 (Takei) digital dynamometers. One hundred and ten outpatients aged >60 years were included. Inter-instrumental reliability was determined. The differences between dynamometers were evaluated by Bland–Altman plots and measurement error. The measurements with Jamar+, and Takei dynamometers were reliable and valid regarding the Jamar dynamometer. Takei and Jamar+ dynamometers overestimated GS over the Jamar dynamometer. Though the differences in the measured values might be disregarded in clinical practice, individuals defined to have low GS varied by the use of different dynamometers. Grip strength better be measured by the same dynamometer in serial measurements of older individuals.
CITATION STYLE
Savas, S., Kilavuz, A., Kayhan Koçak, F. Ö., & Cavdar, S. (2023). Comparison of Grip Strength Measurements by Widely Used Three Dynamometers in Outpatients Aged 60 Years and Over. Journal of Clinical Medicine, 12(13). https://doi.org/10.3390/jcm12134260