The authors assessed the analytical performance of the co-immobilized hexokinase (EC 2.7.1.1) and glucose-6-phosphate dehydrogenase (EC 1.1.1.49) method for D-glucose analysis on the Technicon SMAC. The enzyme-containing coils were usable for one month, or 12000 tests. Bilirubin, hemoglobin, lipemia, creatinine, uric acid, citric acid, and ascorbic acid did not interfere. Results with this method were compared to those by the National Glucose Reference Method. The upper limits of the total error estimate (a combination of random and systematic errors) were 76, 74, and 125 mg/liter at concentrations of 500, 1200, and 3000 mg/liter, respectively. The error estimates were less than allowable errors based on medical usefulness; thus the method was judged to perform acceptably with respect to the Reference Method. The authors also present performance data for the routine SMAC glucose oxidase (EC 1.1.3.4)/peroxidase (EC 1.11.1.7) 3-methyl-2-benzothianolinone hydrazone-N,N-dimethylanilline method, the direct hexokinase method with the Du Pont aca, and the glucose oxidase oxygen-rate method with the Beckman Glucose Analyzer.
CITATION STYLE
Garber, C. C., Feldbruegge, D., Miller, R. C., & Carey, R. N. (1978). Evaluation of the co-immobilized hexokinase/glucose-6-phosphate dehydrogenase method for glucose, as adapted to the Technicon SMAC. Clinical Chemistry, 24(7), 1186–1190. https://doi.org/10.1093/clinchem/24.7.1186