Objective: To describe (1) the radiographic appearance of intact hydraulic urethral sphincters (HUS) and (2) the success of leak detection using clinically feasible methods. Study Design: Prospective, blinded in vitro study. Sample Population: Thirty HUS devices (10 each of 8, 10, and 12mm diameter sizes). Materials and Methods: All devices were inflated with saline (0.9% NaCl) solution to complete occlusion, inspected, and weighed over a 24-hour period for manufacturing defects. HUS phantoms were created to mimic surrounding soft tissues. One randomly selected HUS of each size was evaluated radiographically at different inflation volumes and angles. All HUS systems were then evaluated in random order before and after puncture with volumetry, manometry, radiography, and contrast fluoroscopy. Volumetry was the total volume (mL) retrieved from each HUS system. Manometry was the pressure (cm H2O) within each HUS system. The HUS devices were filled to a known volume before each measurement. Results: When all HUS sizes were considered, volumetry did not reveal significant differences before and after puncture, but manometry was significantly different (P
CITATION STYLE
Tong, K., Nelson, L. L., Hauptman, J., & Nelson, N. C. (2013). Evaluation of percutaneously adjustable hydraulic urethral sphincters with and without induced mechanical failure. Veterinary Surgery, 42(7), 774–782. https://doi.org/10.1111/j.1532-950X.2013.12052.x