Evaluation of a new device for the intraoperative assessment of coronary artery bypasses grafting

2Citations
Citations of this article
14Readers
Mendeley users who have this article in their library.

Abstract

Purpose: The aim of this study was to evaluate a new flow analyze device that can intraoperatively measure compliance and resistance of coronary artery bypass grafting (CABG) by the parameter identification method. Methods: Subjects were 95 patients who underwent CABG. Angiography soon after surgery confirmed patency in 90 grafts and graft failure in 5 grafts. Variables of Aortic pressure and the graft flow were entered into this new device that includes a mathematical coronary circulation model to extrapolate other information the data; thus, we can estimate intraoperative flow rate, resistance, and compliance. Flow rate, resistance, and compliance were compared between patent and failed graft groups and among three different type grafts. Results: There were no differences in flow rate, resistance, and compliance among the three grafts types. Between the patent and failed graft groups, there were statistically significant differences in flow rate and resistance, and compliance tended to be lower for the failed graft group. Conclusions: By measuring graft resistance and compliance, this new device is useful for evaluating graft performance during CABG. © 2011 The Editorial Committee of Annals of Thoracic and Cardiovascular Surgery.

Cite

CITATION STYLE

APA

Nakata, K., Sankai, Y., Akiyama, K., Orime, Y., Kashiwazaki, S., Koba, H., & Shiono, M. (2011). Evaluation of a new device for the intraoperative assessment of coronary artery bypasses grafting. Annals of Thoracic and Cardiovascular Surgery, 17(2), 160–165. https://doi.org/10.5761/atcs.oa.09.01507

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free