Clinical safety of 3-T brain magnetic resonance imaging in newborns

5Citations
Citations of this article
18Readers
Mendeley users who have this article in their library.
Get full text

Abstract

Background: The effects and potential hazards of brain magnetic resonance imaging (MRI) at 3 T in newborns are debated. Objective: Assess the impact of 3-T MRI in newborns on body temperature and physiological parameters. Material and methods: Forty-nine newborns, born preterm and at term, underwent 3-T brain MRI at term-corrected age. Rectal and skin temperature, oxygen saturation and heart rate were recorded before, during and after the scan. Results: A statistically significant increase in skin temperature of 0.6 °C was observed at the end of the MRI scan (P<0.01). There was no significant changes in rectal temperature, heart rate or oxygen saturation. Conclusion: Core temperature, heart rate and oxygen saturation in newborns were not affected by 3-T brain MR scanning.

Cite

CITATION STYLE

APA

Fumagalli, M., Cinnante, C. M., Calloni, S. F., Sorrentino, G., Gorla, I., Plevani, L., … Triulzi, F. (2018). Clinical safety of 3-T brain magnetic resonance imaging in newborns. Pediatric Radiology, 48(7), 992–998. https://doi.org/10.1007/s00247-018-4105-0