Packing performance of helical Guglielmi detachable coil (GDC) 18 in intracranial aneurysms: A comparison with helical GDC 10 coils and complex Trufill/Orbit coils

18Citations
Citations of this article
16Readers
Mendeley users who have this article in their library.

Abstract

BACKGROUND AND PURPOSE: The purpose of this study was to retrospectively compare the packing performance of helical Guglielmi detachable coil (GDC) 18 (thickness, 0.0135-0.015 inch) with the packing performance of both helical GDC 10 (thickness, 0.010 inch) and complex Trufill/Orbit coils (thickness, 0.012 inch). MATERIALS AND METHODS: From our data base, we selected aneurysms that were exclusively coiled with GDC 18 coils. For every aneurysm treated with GDC 18 coils, we tried to identify a volume-matched control aneurysm treated with exclusively GDC 10 coils or exclusively Trufill/Orbit coils. This process resulted in 32 aneurysm pairs treated with either GDC 18 or GDC 10 coils and 35 aneurysm pairs treated with either GDC 18 or Trufill/Orbit coils. RESULTS: The mean packing of 24.2% of aneurysms treated with GDC 18 was significantly higher than the mean packing of 18.3% of aneurysms treated with GDC 10 (P

Cite

CITATION STYLE

APA

Sluzewski, M., & Van Rooij, W. J. (2007). Packing performance of helical Guglielmi detachable coil (GDC) 18 in intracranial aneurysms: A comparison with helical GDC 10 coils and complex Trufill/Orbit coils. American Journal of Neuroradiology, 28(7), 1384–1387. https://doi.org/10.3174/ajnr.A0557