Background: Traditionally, specific ventilators have been manufactured to only provide neonatal mechanical ventilation. However, many of the current generation of ICU ventilators also include a neonatal mode. Methods: Using the IngMar ASL5000 lung simulator the Puritan Bennett 840, the Maquet Servo i, the Viasys AVEA, the GE Engström, the Drager Evita XL and Babylog 8000 Plus were evaluated during assisted ventilation in the pressure assist/control mode. Three lung mechanics were set: resistance 50 cmH2O/L/s, compliance 2 mL/cmH2O; resistance 100 cmH2O/L/s, compliance 1 mL/cmH2O; and resistance 150 cmH2O/L/s, compliance 0.5 mL/cmH2O. A maximum negative pressure drop of 4 and 7 cmH 2O was achieved during simulated inspirations. Each ventilator was evaluated with PEEP 5 cmH2O, peak pressure 20 cmH2O and inspiratory time 0.3 s and with PEEP 10 cmH2O, peak pressure 30 cmH2O and inspiratory time 0.4 s. Each ventilator setting was then repeated with a leak of 0.3 L/min at a constant pressure of 5 cmH2O. Results: Overall each of the 5 ICU ventilators responded faster or greater than the Babylog with respect to: pressure to trigger (except the Servo i), time to trigger (except the Evita XL), time between trigger and return of pressure to baseline, time from start of breath to 90% of peak pressure (except the Avea) and pressure time product of breath activation. Expiratory tidal volume was also greater with all ICU ventilators except the Avea. Variation in mechanics, leak, PEEP and muscular effort had little effect on these differences. Conclusion: All ICU ventilators tested were able to at least equal the performance of the Babylog 8000 Plus on all variables evaluated. © 2008 Springer-Verlag.
CITATION STYLE
Marchese, A. D., Chipman, D., De La Oliva, P., & Kacmarek, R. M. (2009). Adult ICU ventilators to provide neonatal ventilation: A lung simulator study. Intensive Care Medicine, 35(4), 631–638. https://doi.org/10.1007/s00134-008-1332-0
Mendeley helps you to discover research relevant for your work.