The present study investigated whether magnetocardiograms (MCGs) could noninvasively detect the His potential. In 22 patients, the His-ventricular (HV) intervals in MCGs (64-channel system, 2-min signal averaging, filter: 0.1-100 Hz, and band-eliminating filter of 50 Hz) were compared with those recorded in an electrophysiologic study (EPS, filter: 30-400Hz). In 14 of the patients (64 %), the His potential was recorded in the MCGs. There was a correlation between the HV intervals in the MCGs and those in the EPS (R=0.81, p<0.01). This study indicates that MCGs can be used to investigate the specialized conduction systems for which EPS is currently used.
CITATION STYLE
Yamada, S., Kuga, K., On, K., & Yamaguchi, I. (2003). Noninvasive recording of his potential using magnetocardiograms. Circulation Journal, 67(7), 622–624. https://doi.org/10.1253/circj.67.622
Mendeley helps you to discover research relevant for your work.