Absolute measurement of the refractive index of water by a mode-locked laser at 518 nm

10Citations
Citations of this article
10Readers
Mendeley users who have this article in their library.

Abstract

In this paper, we demonstrate a method using a frequency comb, which can precisely measure the refractive index of water. We have developed a simple system, in which a Michelson interferometer is placed into a quartz-glass container with a low expansion coefficient, and for which compensation of the thermal expansion of the water container is not required. By scanning a mirror on a moving stage, a pair of cross-correlation patterns can be generated. We can obtain the length information via these cross-correlation patterns, with or without water in the container. The refractive index of water can be measured by the resulting lengths. Long-term experimental results show that our method can measure the refractive index of water with a high degree of accuracy—measurement uncertainty at 10–5 level has been achieved, compared with the values calculated by the empirical formula.

Cite

CITATION STYLE

APA

Meng, Z., Zhai, X., Wei, J., Wang, Z., & Wu, H. (2018). Absolute measurement of the refractive index of water by a mode-locked laser at 518 nm. Sensors (Switzerland), 18(4). https://doi.org/10.3390/s18041143