Pitfalls and premature failure of the Freedom SOLO stentless valve

22Citations
Citations of this article
31Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

OBJECTIVES: This study reports a series of pitfalls, premature failures and explantations of the third-generation Freedom SOLO (FS) bovine pericardial stentless valve. METHODS: A total of 149 patients underwent aortic valve replacement using the FS. Follow-up was 100% complete with an average observation time of 5.5 ± 2.3 years (maximum 8.7 years) and a total of 825 patient-years. Following intraoperative documentation, all explanted valve prostheses underwent histological examination. RESULTS: Freedom from structural valve deterioration (SVD) at 5, 6, 7, 8 and 9 years was 92, 88, 80, 70 and 62%, respectively. Fourteen prostheses required explantation due to valve-independent dysfunction (n = 5; i.e. thrombus formation, oversizing, aortic dilatation, endocarditis and suture dehiscence) or valve-dependent failure (acute leaflet tears, n = 4 and severe stenosis, n = 5). Thus, freedom from explantation at 5, 6, 7, 8 and 9 years was 95, 94, 91, 81 and 72%, respectively. An acute vertical tear along the non-coronary/right coronary commissure to the base occurred at a mean of 6.0 years (range 4.3-7.3 years) and affected size 25 and 27 prostheses exclusively. Four FS required explantation after a mean of 7.5 years (range 7.0-8.3 years) due to severe functional stenosis and gross calcification that included the entire aortic root. CONCLUSIONS: The FS stentless valve is safe to implant and shows satisfying mid-term results in our single institution experience. Freedom from SVD and explantation decreased markedly after only 6-7 years, so that patients with FS require close observation and follow-up. Exact sizing, symmetric positioning and observing patient limitations are crucial for optimal outcome.

Cite

CITATION STYLE

APA

Stanger, O., Bleuel, I., Reineke, S., Banz, Y., Erdoes, G., Tevaearai, H., … Englberger, L. (2015). Pitfalls and premature failure of the Freedom SOLO stentless valve. European Journal of Cardio-Thoracic Surgery, 48(4), 562–570. https://doi.org/10.1093/ejcts/ezu473

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free