To prevent periprosthetic osteolysis and subsequent aseptic loosening of artificial hip joints, we recently developed a novel acetabular highly cross-linked polyethylene (CLPE) liner with graft polymerization of 2-methacryloyloxyethyl phosphorylcholine (MPC) on its surface. We investigated the wear resistance of the poly(MPC) (PMPC)-grafted CLPE liner during 20 million cycles in a hip joint simulator. We extended the simulator test of one liner to 70 million cycles to investigate the long-term durability of the grafting. Gravimetric, surface, and wear particle analyses revealed that PMPC grafting onto the CLPE liner surface markedly decreased the production of wear particles and showed that the effect of PMPC grafting was maintained through 70 million cycles. We believe that PMPC grafting can significantly improve the wear resistance of artificial hip joints. © 2013 Orthopaedic Research Society. Published by Wiley Periodicals, Inc. J Orthop Res 32:369-376, 2014. © 2013 Orthopaedic Research Society. Published by Wiley Periodicals, Inc.

Cite

CITATION STYLE

APA

Moro, T., Takatori, Y., Kyomoto, M., Ishihara, K., Hashimoto, M., Ito, H., … Kawaguchi, H. (2014). Long-term hip simulator testing of the artificial hip joint bearing surface grafted with biocompatible phospholipid polymer. Journal of Orthopaedic Research, 32(3), 369–376. https://doi.org/10.1002/jor.22526

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free