A 65-year-old woman has a permanent pacemaker implanted on the previous day for episodes of AV block resulting in syncope. The device is a dual chamber Medtronic EnRhythm DR. The atrial sensing at implant was 3 mV, the threshold was 0.8 V at a pulse width of 0.5 ms, and the impedance was 556 Ω. The ventricular sensing at implant was 15 mV, the threshold was 0.4 V at a pulse width of 0.5 ms, and the impedance was 840 Ω. © Springer-Verlag London Limited 2011.