Radiation isocenter is more important part of quality assurance for the linear accelerator (Linac) due to radiation isocenter is a main location in irradiation radiotherapy, isocenter can shift when the gantry and collimator rotation. In general, the radiation isocenter verification using a special film. This research was conducted radiation isocenter verification using computed radiography with digital image processing techniques. Image acquisition was done using the modalities of Linac 6 MV with star shot method is star-shaped beam due to rotation of the collimator, gantry and couch. Then do the delineation on each beam to determine the centroid and beam diameter. By the results of verification of radiation isocenter performed on collimator and the couch, it shows that the size diameter for rotational collimator is 0.632 mm and 0.458 mm for the couch. Based on AAPM report 40 about the size of the Linac radiation isocenter diameter used in this study is still in good condition and worth to be operated because the value of the radiation isocenter diameter is below 2 mm.
CITATION STYLE
Irsal, M., Hidayanto, E., & Sutanto, H. (2017). Verification of Radiation Isocenter on Linac Beam 6 MV using Computed Radiography. In Journal of Physics: Conference Series (Vol. 855). Institute of Physics Publishing. https://doi.org/10.1088/1742-6596/855/1/012020
Mendeley helps you to discover research relevant for your work.