Spectrophotometry assays to determine G6PD activity from Trinity Biotech and Pointe Scientific G6PD show good correlation

17Citations
Citations of this article
22Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Objectives: Spectrophotometry kits from Pointe Scientific (PS; USA) were compared to kits from Trinity Biotech (Trinity; Ireland) in 50 venous blood samples from purposively selected individuals in Bangladesh. Repeatability and inter-assay variability were assessed by Students t-test, Bland-Altman plot and Pearson correlation coefficient (r). The median glucose-6-phosphate dehydrogenase (G6PD) activity of all G6PD normal participants was calculated per assay and defined as 100% activity. Performance was calculated considering 30% and 70% cut off activities and Trinity as reference. Results: The intra-assay correlation of Trinity (r = 0.9841, p < 0.001) and PS (r = 0.9833, p < 0.001) did not differ significantly (p = 0.904). Both assays were closely correlated (r = 0.9799, p < 0.001), with a mean difference of 0.1 U/gHb (95% limit of agreement: - 1.32 to 1.57). At 30% cut off PS had a sensitivity of 100% (95% confidence interval (95 CI) 59.0-100.0) and specificity of 100% (95% CI 91.8 to 100.0), at 70% cut-off of 100% (95% CI 79.4-100.0) and 97.1% (95% CI 84.7-99.9) respectively. The G6PD assay from PS is a reliable alternative to the assay from Trinity.

Author supplied keywords

Cite

CITATION STYLE

APA

Alam, M. S., Kibria, M. G., Jahan, N., Price, R. N., & Ley, B. (2018). Spectrophotometry assays to determine G6PD activity from Trinity Biotech and Pointe Scientific G6PD show good correlation. BMC Research Notes, 11(1). https://doi.org/10.1186/s13104-018-3964-7

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free