Development and validation of a high-performance liquid chromatography method for determination of cefquinome concentrations in sheep plasma and its application to pharmacokinetic studies

54Citations
Citations of this article
50Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Cefquinome has a broad spectrum of antibacterial activity and was developed especially for use in animals. A simple and sensitive high-performance liquid chromatography (HPLC) method with UV-visible detection for quantification of cefquinome concentrations in sheep plasma was developed and validated. Separation of cefquinome from plasma components was achieved on a Phenomenex Gemini C 18 column (250 mm by 4.6 mm; internal diameter [i.d.], 5 μm). The mobile phase consisted of acetonitrile and 0.1% trifluoroacetic acid in water and was delivered at a rate of 0.9 ml/min. A simple and rapid sample preparation involved the addition of methanol to 200 μl of plasma to precipitate plasma proteins followed by direct injection of 50 μl of supernatant into the high-performance liquid chromatography system. The linearity range of the proposed method was 0.02 to 12 μg/ml. The intraday and interday coefficients of variation obtained from cefquinome were less than 5%, and biases ranged from -3.76% to 1.24%. Mean recovery based on low-, medium-, and high-quality control standards ranged between 92.0 and 93.9%. Plasma samples were found to be stable in various storage conditions (freeze-thaw, postpreparative, short-term, and long-term stability). The method described was found to be readily available, practicable, cheap, rapid, sensitive, precise, and accurate. It was successfully applied to the study of the pharmacokinetics of cefquinome in sheep. This method can be very useful and an alternate to performing pharmacokinetic studies in the determination of cefquinome for clinical use. Copyright © 2011, American Society for Microbiology. All Rights Reserved.

Cite

CITATION STYLE

APA

Uney, K., Altan, F., & Elmas, M. (2011). Development and validation of a high-performance liquid chromatography method for determination of cefquinome concentrations in sheep plasma and its application to pharmacokinetic studies. Antimicrobial Agents and Chemotherapy, 55(2), 854–859. https://doi.org/10.1128/AAC.01126-10

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free