Determination of reduced and oxidized coenzyme Q10 in canine plasma and heart tissue by HPLC-ECD: Comparison with LC-MS/MS quantification

17Citations
Citations of this article
36Readers
Mendeley users who have this article in their library.

Abstract

Coenzyme Q10 (Q10) plays an important role in mammals for energy production in the mitochondria, and as a potent antioxidant. Oxidation ratio (% oxidized in relation to total Q10) has been proposed as an important biomarker. A sensitive and reproducible HPLC-ECD method was developed for determination of reduced and oxidized Q10 in canine plasma and heart tissue. Chromatographic separation was achieved in 10 min using a Waters Nova-pak C18 column and a mobile phase with lithium perchlorate in ethanol/methanol/2-propanol. The validation showed satisfying results. Excellent linear correlation was found (r2 > 0.9997), intra-and inter-day precisions were below 6.5% (n = 5) and recoveries were between 89 and 109% (n = 5). Sensitivity stated as Lower Limit of Quantification (LLOQ) was 10 nM. Acceptable stability of both extracted and un-extracted samples was observed. The plasma concentration range of total Q10 was found to be between 0.64 and 1.24 µg/mL. Comparison with a developed LC-MS/MS method showed a correlation of r = 0.85 for reduced Q10 and r = 0.60 for oxidized Q10 (N = 17). However, average results were around 30% lower for ubiquinol using the LC-MS/MS method as compared with the HPLC-ECD analysis. The two methods are therefore not considered to be interchangeable.

Cite

CITATION STYLE

APA

Schou-Pedersen, A. M. V., Schemeth, D., & Lykkesfeldt, J. (2019). Determination of reduced and oxidized coenzyme Q10 in canine plasma and heart tissue by HPLC-ECD: Comparison with LC-MS/MS quantification. Antioxidants, 8(8). https://doi.org/10.3390/antiox8080253

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free