Improvement of spiral MRI with the measured k-space trajectory

13Citations
Citations of this article
24Readers
Mendeley users who have this article in their library.

Abstract

The k-space trajectory of a spiral imaging sequence was measured with a self-encoding technique. The image quality improved dramatically when reconstructed with the measured k-space trajectory. There were substantial artifacts in images reconstructed with the derived k-space trajectory under the assumption of gradient system linearity. The results indicated the nonlinearity of the gradient system and the effectiveness of the correction technique.

Cite

CITATION STYLE

APA

Ding, X., Tkach, J., Ruggieri, P., Perl, J., & Masaryk, T. (1997). Improvement of spiral MRI with the measured k-space trajectory. Journal of Magnetic Resonance Imaging, 7(5), 938–940. https://doi.org/10.1002/jmri.1880070525