Open-Source Tracked Ultrasound with Anser Electromagnetic Tracking

1Citations
Citations of this article
6Readers
Mendeley users who have this article in their library.
Get full text

Abstract

Image-guided interventions (IGT) have shown a huge potential to improve medical procedures or even allow for new treatment options. Most ultrasound(US)-based IGT systems use electromagnetic (EM) tracking for localizing US probes and instruments. However, EM tracking is not always reliable in clinical settings because the EM field can be disturbed by medical equipment. So far, most researchers used and studied commercial EM trackers with their IGT systems which in turn limited the possibilities to customize the trackers in order minimize distortions and make the systems robust for clinical use. In light of current good scientific practice initiatives that increasingly request research to publish the source code corresponding to a paper, the aim of this work was to test the feasibility of using the open-source EM tracker (Anser EMT) for localizing US probes in a clinical US suite for the first time. The standardized protocol of Hummel et al. yielded a jitter of 0.1 � 0.1mm and a position error of 1.1�0.7mm, which is comparable to 0.1mm and 1.0mm of a commercial NDI Aurora system. The rotation error of Anser EMT was 0.15�0.16, which is lower than at least 0.4 for the commercial tracker. We consider tracked US as feasible with Anser EMT if an accuracy of 1�2mm is sufficient for a specific application.

Cite

CITATION STYLE

APA

Franz, A. M., Jaeger, H. A., Seitel, A., Cantillon-Murphy, P., & Maier-Hein, L. (2019). Open-Source Tracked Ultrasound with Anser Electromagnetic Tracking. In Informatik aktuell (pp. 232–237). Springer Berlin Heidelberg. https://doi.org/10.1007/978-3-658-25326-4_52

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free