Point-of-care measurement of activated clotting time for cardiac surgery as measured by the Hemochron signature elite and the Abbott i-STAT: Agreement, concordance, and clinical reliability

19Citations
Citations of this article
52Readers
Mendeley users who have this article in their library.
Get full text

Abstract

Background: Since inadequate heparin anticoagulation and insufficient reversal can result in complications during cardiopulmonary bypass (CPB) surgery, heparin anticoagulation monitoring by point-of-care (POC) activated clotting time (ACT) measurements is essential for CPB initiation, maintainance, and anticoagulant reversal. However, concerns exist regarding reproducibility of ACT assays and comparability of devices. Methods: We evaluated the agreement of ACT assays using four parallel measurements performed on two commonly used devices each (i.e., two Hemochron Signature Elite (Hemochron) and two Abbott i-STAT (i-STAT) devices, respectively). Blood samples from 30 patients undergoing cardiac surgery on CPB were assayed at specified steps (baseline, after heparin administration, after protamine administration) with four parallel measurements (two of each device type) using commercial Kaolin activated assays provided by the respective manufactures. Measurements were compared between identical and different device types using linear regression, Bland-Altman analyses, and calculation of Cohen's kappa coefficient. Results: Parallel i-STAT ACTs demonstrated a good linear correlation (r = 0.985). Bias, as determined by Bland-Altman analysis, was low (- 3.8 s; 95% limits of agreement (LOA): - 77.8 -70.2 s), and Cohen's Kappa demonstrated good agreement (kappa = 0.809). Hemochron derived ACTs demonstrated worse linear correlation (r = 0.782), larger bias with considerably broader LOA (- 13.14 s; 95%LOA:-316.3-290 s), and lesser concordance between parallel assays (kappa = 0.554). Although demonstrating a fair linear correlation (r = 0.815), parallel measurements on different ACT-devices showed large bias (-20s; 95% LOA: - 290-250 s) and little concordance (kappa = 0.368). Overall, disconcordant results according to clinically predefined target values were more frequent with the Hemochron than i-STAT. Furthermore, while discrepancies in ACT between two parallel iSTAT assays showed little or no clinical relevance, deviations from parallel Hemochron assays and iSTAT versus Hemochron measurements revealed marked and sometimes clinically critical deviations. Conclusion: Currently used ACT point-of-care devices cannot be used interchangeably. Furthermore, our data question the reliability of the Hemochron in assessing adequacy of heparin anticoagulation monitoring for CPB.

References Powered by Scopus

Activated Coagulation Time of Whole Blood

392Citations
N/AReaders
Get full text

2017 EACTS/EACTA Guidelines on patient blood management for adult cardiac surgery: The task force on patient blood management for adult cardiac surgery of the European Association for Cardio-Thoracic Surgery (EACTS) and the European Association of Cardiothoracic Anaesthesiology (EACTA)

315Citations
N/AReaders
Get full text

Comparison of thromboelastometry (ROTEM<sup>®</sup>) with standard plasmatic coagulation testing in paediatric surgery

229Citations
N/AReaders
Get full text

Cited by Powered by Scopus

Managing the coagulopathy associated with cardiopulmonary bypass

83Citations
N/AReaders
Get full text

Activated Clotting Time (ACT) for Monitoring of Low-Dose Heparin: Performance Characteristics in Healthy Adults and Critically Ill Patients

12Citations
N/AReaders
Get full text

Running after activated clotting time values in patients receiving direct oral anticoagulants: A potentially dangerous race. results from a prospective study in atrial fibrillation catheter ablation procedures

9Citations
N/AReaders
Get full text

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Dirkmann, D., Nagy, E., Britten, M. W., & Peters, J. (2019). Point-of-care measurement of activated clotting time for cardiac surgery as measured by the Hemochron signature elite and the Abbott i-STAT: Agreement, concordance, and clinical reliability. BMC Anesthesiology, 19(1). https://doi.org/10.1186/s12871-019-0846-z

Readers' Seniority

Tooltip

PhD / Post grad / Masters / Doc 16

67%

Researcher 6

25%

Professor / Associate Prof. 2

8%

Readers' Discipline

Tooltip

Medicine and Dentistry 19

73%

Nursing and Health Professions 3

12%

Engineering 3

12%

Pharmacology, Toxicology and Pharmaceut... 1

4%

Article Metrics

Tooltip
Social Media
Shares, Likes & Comments: 11

Save time finding and organizing research with Mendeley

Sign up for free