Effect of radiant warmer on transepidermal water loss (TEWL) and skin hydration in preterm infants

11Citations
Citations of this article
29Readers
Mendeley users who have this article in their library.
Get full text

Abstract

Objective: To evaluate the effect of radiant warmers on skin barrier function in preterm infants. Methodology: Transepidermal water loss (TEWL) and stratum corneum hydration were measured in 30 preterm infants (birth weight 825 to 2220 g) in seven body areas: forehead, upper back, cubital fossa, palms, soles. abdomen, and inguinal region. Measurements were performed under radiant warmer and incubator conditions. Each patient served as his/her control. Results: TEWL was significantly higher in the radiant warmer compared to the incubator condition in only two areas: forehead and back. The overall mean difference in percentage TEWL between the conditions Stratum corneum hydration was not affected by the radiant warmer. Conclusions: The use of radiant warmers does not significantly decrease barrier function in the preterm infant. © 2004 Nature Publishing Group. All rights reserved.

Cite

CITATION STYLE

APA

Maayan-Metzger, A., Yosipovitch, G., Hadad, E., & Sirota, L. (2004). Effect of radiant warmer on transepidermal water loss (TEWL) and skin hydration in preterm infants. Journal of Perinatology, 24(6), 372–375. https://doi.org/10.1038/sj.jp.7211107