Spatially uniform reconstitution of dried reagents is critical to the function of paper microfluidic devices. Advancing fluid fronts in paper microfluidic devices drive (convect) and concentrate rehydrated reagents to the edges, causing steep chemical gradients and imperfect mixing. This largely unsolved problem in paper microfluidics is exacerbated by increasing device dimensions. In this article, we demonstrate that mixing of dried reagents with a rehydrating fluid in paper microfluidics may be significantly enhanced by stacking paper layers having different wicking rates. Compared to single-layer paper membranes, stacking reduced the “non-reactive area”, i.e. area in which the reconstituted reagents did not interact with the rehydrating fluid, by as much as 97% in large (8 cm × 2 cm) paper membranes. A paper stack was designed to collect ~0.9 ml liquid sample and uniformly mix it with dried reagents. Applications of this technology are demonstrated in two areas: (i) collection and dry storage of sputum samples for tuberculosis testing, and (ii) salivary glucose detection using an enzymatic assay and colorimetric readout. Maximizing the interaction of liquids with dried reagents is central to enhancing the performance of all paper microfluidic devices; this technique is therefore likely to find important applications in paper microfluidics.

Cite

CITATION STYLE

APA

Das, D., Dsouza, A., Kaur, N., Soni, S., & Toley, B. J. (2019). Paper Stacks for Uniform Rehydration of Dried Reagents in Paper Microfluidic Devices. Scientific Reports, 9(1). https://doi.org/10.1038/s41598-019-52202-9

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free