Significant improvements have been made in Positron Emission Tomography (PET) to enhance the image quality, namely, the development of time-of-flight (TOF) technology. This technique is useful to localize the emission point of the beta plus-emitter (β+) radiopharmaceutical inside the body, allowing better lesion contrast and leading to a short scan time. The main goal of this study is to investigate the shortest acquisition time without compromising the image quality in both a NEMA body phantom and patients, using a TOF PET/Computed Tomography (PET/CT) scanner and the radionuclide Gallium-68 (68Ga). Image quality parameters and quantification in terms of standardized uptake value (SUV) were acquired. A time between 45 and 60 s per bed position is proposed for future clinical practices.
CITATION STYLE
Oliveira, J., Parafita, R., & Branco, S. (2018). Minimisation of acquisition time in a TOF PET/CT scanner without compromising image quality. Lecture Notes in Computational Vision and Biomechanics, 27, 27–42. https://doi.org/10.1007/978-3-319-68195-5_4