Minimisation of acquisition time in a TOF PET/CT scanner without compromising image quality

0Citations
Citations of this article
2Readers
Mendeley users who have this article in their library.
Get full text

Abstract

Significant improvements have been made in Positron Emission Tomography (PET) to enhance the image quality, namely, the development of time-of-flight (TOF) technology. This technique is useful to localize the emission point of the beta plus-emitter (β+) radiopharmaceutical inside the body, allowing better lesion contrast and leading to a short scan time. The main goal of this study is to investigate the shortest acquisition time without compromising the image quality in both a NEMA body phantom and patients, using a TOF PET/Computed Tomography (PET/CT) scanner and the radionuclide Gallium-68 (68Ga). Image quality parameters and quantification in terms of standardized uptake value (SUV) were acquired. A time between 45 and 60 s per bed position is proposed for future clinical practices.

Cite

CITATION STYLE

APA

Oliveira, J., Parafita, R., & Branco, S. (2018). Minimisation of acquisition time in a TOF PET/CT scanner without compromising image quality. Lecture Notes in Computational Vision and Biomechanics, 27, 27–42. https://doi.org/10.1007/978-3-319-68195-5_4