Poly-L-lactic acid β-tricalcium phosphate screws: A preliminary in vivo biocompatibility study

12Citations
Citations of this article
11Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

The aim of this study is to assess the biocompatibility of two types of Poly-L-lactic acid (PLLA) screws (with either hydroxiapatite (HA) or β-tricalcium phosphate (β-TCP)) implanted in the left femur of four sheep euthanized at 42, 50, 57 and 84 days after surgery. Titanium screws were also implanted for comparison purposes. No signs of inflammation were seen in the 240 specimens. A rating of "+/-"for macrophages and "-"for neutrophils was assigned to all specimens. All specimens were assigned a rating which ranged from "+/-" to "+++" for fibroblasts and osteoblasts. The presence of macrophages, neutrophils and fibroblasts/osteoblasts was not statistically different for the four implantation periods. PLLA implants with β-TCP have a biocompatibility comparable to PLLA implants with HA. Copyright © by Biolife, s.a.s.

Cite

CITATION STYLE

APA

Magarelli, N., Savastano, M. A., Palmieri, D., Zappacosta, R., Lattanzio, G., Salini, V., … Colosimo, C. (2007). Poly-L-lactic acid β-tricalcium phosphate screws: A preliminary in vivo biocompatibility study. International Journal of Immunopathology and Pharmacology, 20(1), 207–211. https://doi.org/10.1177/039463200702000126