Effect of insulin infusion line on glycaemic variability in a perioperative high dependency unit (HDU): a prospective randomised controlled trial

1Citations
Citations of this article
17Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Background: Glucose control is an important issue in post-operative patients. The objective here was to compare two insulin infusion lines by syringe pumps to assess the impact of medical devices on glycaemic variability in surgical patients under intensive insulin therapy. This open, prospective, single-centre randomised study was conducted in a fifteen-bed perioperative high dependency unit (HDU) in a university hospital. In total, 172 eligible patients receiving insulin therapy agreed to participate in the study. Subcutaneous continuous glucose monitoring was set up for all patients and an optimised system with a dedicated insulin infusion line for half of the patients. Results: Eighty-six patients were infused via the optimised infusion line and 86 patients via the standard infusion line. No significant difference was found according to the glycaemic lability index score [mean difference between groups (95% CI): −0.09 (−0.34; 0.16), p = 0.49 after multiple imputation]. A glucose control monitoring system indicated a trend towards differences in the duration of hypoglycaemia (blood glucose level below 70 mg dl−1 (3.9 mmol l−1) over 1000 h of insulin infusion (9.7 ± 25.0 h in the standard group versus 4.4 ± 14.8 h in the optimised group, p = 0.059) and in the number of patients experiencing at least one hypoglycaemia incident (25.7 vs. 12.9%, p = 0.052). Time in the target range was similar for both groups. Conclusions: The use of optimised infusion line with a dedicated insulin infusion line did not reduce glycaemic variability but minimised the incidence of hypoglycaemia events. The choice of the medical devices used to infuse insulin seems important for improving the safety of insulin infusion in perioperative HDU.

References Powered by Scopus

Intensive insulin therapy in critically ill patients

8596Citations
N/AReaders
Get full text

Intensive versus conventional glucose control in critically Ill patients

4120Citations
N/AReaders
Get full text

Intensive insulin therapy in the medical ICU

3093Citations
N/AReaders
Get full text

Cited by Powered by Scopus

In vitro assessment of the influence of intravenous extension set materials on insulin aspart drug delivery

15Citations
N/AReaders
Get full text

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Genay, S., Décaudin, B., Ethgen, S., Alluin, A., Babol, E., Labreuche, J., … Lebuffe, G. (2017). Effect of insulin infusion line on glycaemic variability in a perioperative high dependency unit (HDU): a prospective randomised controlled trial. Annals of Intensive Care, 7(1). https://doi.org/10.1186/s13613-017-0298-x

Readers' Seniority

Tooltip

PhD / Post grad / Masters / Doc 8

80%

Professor / Associate Prof. 1

10%

Researcher 1

10%

Readers' Discipline

Tooltip

Medicine and Dentistry 6

60%

Nursing and Health Professions 2

20%

Pharmacology, Toxicology and Pharmaceut... 1

10%

Neuroscience 1

10%

Article Metrics

Tooltip
Social Media
Shares, Likes & Comments: 12

Save time finding and organizing research with Mendeley

Sign up for free