Prospective, randomized, multicenter, controlled trial (TRIATHRON 1) on a new antithrombogenic hydrophilic dialysis membrane

16Citations
Citations of this article
21Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Introduction: Hemodialysis treatment requires anticoagulation to prevent thrombosis of the dialyzer. The Hydrolink ® (NV series; Toray) has been designed to reduce thrombotic complications by increasing membrane hydrophilic properties. Previous studies have confirmed reduced platelet activation, improved removal of β2-microglobulin and excellent small-solute removal. Methods: We designed a prospective, multi-centered, randomized clinical study to compare the antithrombogenic effects (platelet count) of NV dialyzers versus conventional treatment. To compare the possibility of performing heparin-free dialysis, we carried out progressive heparin reduction tests. Patients with an average platelet count lower than 170,000 cells/mm3 using standard high flux membranes in the 6 months prior to the study were enrolled and randomized. Patients were either dialyzed for 6 months without changing the previous membrane (control group) or treated with the Hydrolink® membrane (NV group). After the third week, the heparin reduction test was conducted for 5 weeks in order to assess the minimum amount of anticoagulant needed to safely perform a 4-hour dialysis treatment. Performance and safety were evaluated measuring platelet count and activation, middle-molecule removal rate and nutritional status. Results: We found no significant difference in platelet count, platelet activation factors β-thromboglobulin and platelet factor 4 (PF-4), between the groups. More patients in the study group reached heparin-free dialysis without clotting events during the heparin reduction test. The NV dialyzers displayed anti-thrombogenic effects as compared to conventional dialyzers. Conclusions: The NV dialyzer series is safe with no adverse events reported. Further studies are required to understand the mechanisms of anti-thrombogenic effects.

Cite

CITATION STYLE

APA

Ronco, C., Brendolan, A., Nalesso, F., Zanella, M., De Cal, M., Corradi, V., … Alghisi, A. (2017). Prospective, randomized, multicenter, controlled trial (TRIATHRON 1) on a new antithrombogenic hydrophilic dialysis membrane. International Journal of Artificial Organs, 40(5), 234–239. https://doi.org/10.5301/ijao.5000608

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free