Kerma area product (KaP) and scatter measurements for intraoral X-ray machines using three different types of round collimation compared with rectangular beam limiter

8Citations
Citations of this article
26Readers
Mendeley users who have this article in their library.

Abstract

Objectives: The objective of the study was to determine the radiation dose reduction achieved when rectangular collimation was used on various round collimators. In addition, we evaluated the tissue doses imparted to various head and neck organs. Methods: To evaluate the variation in radiation output based on the variable geometric configurations, the kerma area product (KAP) was measured using a commercially available KAP-meter with an internal ion chamber capable of detecting both radiation dose (µGy) and the primary X-ray beam area. The KAP was measured using standard 20.4, 25.7, and 31.7 cm 2 round collimators with and without rectangular X-ray field restrictors. To evaluate the potential change in patient scatter radiation dose, an adult head phantom was loaded with thin strips of gafchromic film. A full mouth X-ray series was acquired with various geometric configurations. The films were quantified using a calibration factor to yield absorbed organ doses for the eyes, thyroid, and salivary glands. Results: With the use of rectangular collimator, the KAP for a 31.7 cm 2 round collimator was reduced by up to 60% while the 20.4 cm 2 round collimator elicited a reduction from up to 40%. In the organ study, results of up to 81% reduction in scatter radiation dose were observed. conclusions: Although, US FDA regulations allow a maximum beam size of 38.5 cm 2 on the patient skin, this study suggests that the use of rectangular collimators provide clinically relevant dose reduction for patients, even when using smaller round collimation, hence the use of rectangular collimation for all intraoral radiographic procedures is highly recommended.

References Powered by Scopus

ICRP publication 118: ICRP Statement on Tissue Reactions and Early and Late Effects of Radiation in Normal Tissues and Organs — Threshold Doses for Tissue Reactions in a Radiation Protection Context

1063Citations
N/AReaders
Get full text

Patient risk related to common dental radiographic examinations: The impact of 2007 International Commission on Radiological Protection recommendations regarding dose calculation

315Citations
N/AReaders
Get full text

Radiobiologic risk estimation from dental radiology. Part I. Absorbed doses to critical organs

67Citations
N/AReaders
Get full text

Cited by Powered by Scopus

Assessment of dental radiation dose reduction utilizing mathematical pediatric phantom models: Applications in clinical practice

3Citations
N/AReaders
Get full text

Radiation dose reduction using novel size 1 and size 0 rectangular collimators in pediatric dental imaging

1Citations
N/AReaders
Get full text

A study for revising diagnostic reference levels for intra-oral radiology in Cyprus

0Citations
N/AReaders
Get full text

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Magill, D., Huu Ngo, N. J., Felice, M. A., & Mupparapu, M. (2019). Kerma area product (KaP) and scatter measurements for intraoral X-ray machines using three different types of round collimation compared with rectangular beam limiter. Dentomaxillofacial Radiology, 48. https://doi.org/10.1259/dmfr.20180183

Readers' Seniority

Tooltip

Researcher 4

40%

PhD / Post grad / Masters / Doc 3

30%

Lecturer / Post doc 2

20%

Professor / Associate Prof. 1

10%

Readers' Discipline

Tooltip

Medicine and Dentistry 6

55%

Nursing and Health Professions 2

18%

Earth and Planetary Sciences 2

18%

Materials Science 1

9%

Save time finding and organizing research with Mendeley

Sign up for free