Heparin anticoagulation monitoring by point-of-care activated clotting time (ACT) is essential for cardiopulmonary bypass (CPB) initiation, maintenance, and anticoagulant reversal. Concerns exist regarding the comparability of kaolin activated ACT devices. The current study, therefore, evaluated the agreement of ACT assays using parallel measurements performed on two commonly used devices. Measurements were conducted in a split-sample fashion on both the ACT Plus (Medtronic, Minneapolis, MN) and i-STAT (Abbott Point of Care, Princeton, NJ) analyzers. Blood samples from 100 adult patients undergoing elective cardiac surgery with CPB were assayed at specified time points: before heparinization, following systemic heparinization, after CPB initiation, every 30 minutes during CPB, and following protamine administration. A cutoff value of 400 seconds (s) was used as part of the local protocol. Measurements were compared using t tests or Wilcoxon signed-rank tests, linear regression, and Bland-Altman analyses. Parallel ACT measurements demonstrated a good linear correlation (r =.831, p
CITATION STYLE
Vranken, N. P. A., Theelen, A. J. A. J., Orbons, T., Barenbrug, P. J. C., & Weerwind, P. W. (2021). Point-of-Care Measurement of Kaolin Activated Clotting Time during Cardiopulmonary Bypass: A Single Sample Comparison between ACT Plus and i-STAT. Journal of Extra-Corporeal Technology, 53(1), 57–61. https://doi.org/10.1051/ject/202153057