Free calcium in serum. I. Determination with the ion specific electrode and factors affecting the results

66Citations
Citations of this article
6Readers
Mendeley users who have this article in their library.
Get full text

Abstract

Free (ionic) calcium in human serum was measured with a commercially available ion specific electrode, together with a saturated KCl salt bridge, a saturated calomel reference electrode, and a direct reading electrometer. Effects of sodium, potassium, and magnesium on the electrode values for calcium were assessed and a rationale demonstrated for the use of calcium standards containing 140 mmol of NaCl, 5 mmol of KCl, and 0.55 mmol of MgCl2 per liter. Trypsin, triethanolamine, and heparin bind calcium; therefore, their use is to be avoided. Variability is greatly decreased if the serum is from a fasting subject and if the collection tube is completely filled, transported on ice, and handled anaerobically. The range of free calcium for 86 apparently healthy adults was 1.175-1.375 mmol/liter when measured at 25°C. The precision (coefficient of variation) of duplicates was 0.6%; the long term precision 2.0%.

Cite

CITATION STYLE

APA

Ladenson, J. H., & Bowers, G. N. (1973). Free calcium in serum. I. Determination with the ion specific electrode and factors affecting the results. Clinical Chemistry, 19(6), 565–574. https://doi.org/10.1093/clinchem/19.6.565