The dosimetric effect of inhomogeneity correction in dynamic conformal arc stereotactic body radiation therapy for lung tumors

12Citations
Citations of this article
28Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

For patients treated with lung stereotactic body radiation therapy (SBRT) using dynamic conformal arcs, the influence of inhomogeneity correction (IC) on normal tissue and tumor dosimetry was studied. For the same numbers of monitor units, the planning target volume equivalent uniform doses calculated without path-length IC were lower than those calculated with IC (mean difference 18%, range 1% to 34%; p < 0.0001). Normal lung dose differences were of the same magnitude in the opposite direction. In reports of SBRT, it will be helpful to maintain clear communication about the type of IC used to avoid future uncertainties about true normal tissue tolerance and tumor dose-response relationships. © 2006 Am. Coll. Med. Phys.

Cite

CITATION STYLE

APA

Kavanagh, B., Ding, M., Schefter, T., Stuhr, K., & Newman, F. (2006). The dosimetric effect of inhomogeneity correction in dynamic conformal arc stereotactic body radiation therapy for lung tumors. Journal of Applied Clinical Medical Physics, 7(2), 58–63. https://doi.org/10.1120/jacmp.v7i2.2236

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free