Integration and calibration of non-dispersive infrared (NDIR) CO2 low-cost sensors and their operation in a sensor network covering Switzerland

46Citations
Citations of this article
67Readers
Mendeley users who have this article in their library.

Abstract

More than 300 non-dispersive infrared (NDIR) CO2 low-cost sensors labelled as LP8 were integrated into sensor units and evaluated for the purpose of long-term operation in the Carbosense CO2 sensor network in Switzerland. Prior to deployment, all sensors were calibrated in a pressure and climate chamber and in ambient conditions co-located with a reference instrument. To investigate their long-term performance and to test different data processing strategies, 18 sensors were deployed at five locations equipped with a reference instrument after calibration. Their accuracy during 19 to 25 months deployment was between 8 and 12ppm. This level of accuracy requires careful sensor calibration prior to deployment, continuous monitoring of the sensors, efficient data filtering, and a procedure to correct drifts and jumps in the sensor signal during operation. High relative humidity (> ∼85%) impairs the LP8 measurements, and corresponding data filtering results in a significant loss during humid conditions. The LP8 sensors are not suitable for the detection of small regional gradients and long-term trends. However, with careful data processing, the sensors are able to resolve CO2 changes and differences with a magnitude larger than about 30ppm. Thereby, the sensor can resolve the site-specific CO2 signal at most locations in Switzerland. A low-power network (LPN) using LoRaWAN allowed for reliable data transmission with low energy consumption and proved to be a key element of the Carbosense low-cost sensor network.

References Powered by Scopus

Can commercial low-cost sensor platforms contribute to air quality monitoring and exposure estimates?

720Citations
N/AReaders
Get full text

A machine learning calibration model using random forests to improve sensor performance for lower-cost air quality monitoring

373Citations
N/AReaders
Get full text

Field calibration of a cluster of low-cost commercially available sensors for air quality monitoring. Part B: NO, CO and CO<inf>2</inf>

273Citations
N/AReaders
Get full text

Cited by Powered by Scopus

Observed decreases in on-road CO2 concentrations in Beijing during COVID-19 restrictions

26Citations
N/AReaders
Get full text

Carbon dioxide sensing—biomedical applications to human subjects

25Citations
N/AReaders
Get full text

The Berkeley Environmental Air-quality and CO2 Network: Field calibrations of sensor temperature dependence and assessment of network scale CO2 accuracy

20Citations
N/AReaders
Get full text

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Müller, M., Graf, P., Meyer, J., Pentina, A., Brunner, D., Perez-Cruz, F., … Emmenegger, L. (2020). Integration and calibration of non-dispersive infrared (NDIR) CO2 low-cost sensors and their operation in a sensor network covering Switzerland. Atmospheric Measurement Techniques, 13(7), 3815–3834. https://doi.org/10.5194/amt-13-3815-2020

Readers' Seniority

Tooltip

PhD / Post grad / Masters / Doc 15

54%

Researcher 10

36%

Professor / Associate Prof. 2

7%

Lecturer / Post doc 1

4%

Readers' Discipline

Tooltip

Environmental Science 8

36%

Engineering 7

32%

Earth and Planetary Sciences 4

18%

Agricultural and Biological Sciences 3

14%

Save time finding and organizing research with Mendeley

Sign up for free