Human Factors Usability and Validation Studies of a Glucagon Autoinjector in a Simulated Severe Hypoglycemia Rescue Situation

39Citations
Citations of this article
65Readers
Mendeley users who have this article in their library.

Abstract

Background: A room-temperature stable, soluble liquid glucagon formulation loaded into a prefilled, single-use, two-step autoinjector is under development for severe hypoglycemia rescue. We report a human factors validation program evaluating the glucagon autoinjector (GAI) (Gvoke HypoPen™; Xeris Pharmaceuticals, Inc., Chicago, IL) versus marketed glucagon emergency kits (GEKs) for managing severe hypoglycemia. Methods: A simulated-use human factors usability study was conducted with the GAI versus marketed GEKs in 16 participants, including adult caregivers and first responders, experienced with glucagon administration. A summative human factors validation study of the GAI was conducted with 75 volunteers. Participants were (1) trained on the device and procedure or (2) given time to individually read the instructions and familiarize themselves with the device. Participants returned a week later to perform an unaided rescue attempt that simulated rescue of patients with diabetes suffering a hypoglycemia emergency. Participant actions were recorded for critical rescue tasks and use errors. Results: In the usability study, 88% (14) successfully administered a rescue injection using the GAI versus 31% (5) using GEKs (P < 0.05). Mean total rescue time of use was 47.9 s with the GAI versus 109.0 s with GEKs (P < 0.05). In the validation study, 98.7% successfully administered the rescue injection using the GAI. Overall, there were no patterns of differences between trained versus untrained participants, between caregivers versus first responders or between adults versus adolescents. Conclusion: The GAI and instructional materials can be correctly, safely, and effectively used by intended user, which support continued development of the GAI as an alternative to GEKs.

References Powered by Scopus

Increased mortality of patients with diabetes reporting severe hypoglycemia

469Citations
N/AReaders
Get full text

A critical review of the literature on fear of hypoglycemia in diabetes: Implications for diabetes management and patient education

460Citations
N/AReaders
Get full text

Hypoglycaemia in Type 2 diabetes

424Citations
N/AReaders
Get full text

Cited by Powered by Scopus

9. Pharmacologic Approaches to Glycemic Treatment: Standards of Care in Diabetes—2024

0
320Citations
N/AReaders
Get full text

Management of Individuals With Diabetes at High Risk for Hypoglycemia: An Endocrine Society Clinical Practice Guideline

43Citations
N/AReaders
Get full text

Dasiglucagon-a next-generation glucagon analog for rapid and effective treatment of severe hypoglycemia: Results of phase 3 randomized double-blind clinical trial

43Citations
N/AReaders
Get full text

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Valentine, V., Newswanger, B., Prestrelski, S., Andre, A. D., & Garibaldi, M. (2019). Human Factors Usability and Validation Studies of a Glucagon Autoinjector in a Simulated Severe Hypoglycemia Rescue Situation. Diabetes Technology and Therapeutics, 21(9), 522–530. https://doi.org/10.1089/dia.2019.0148

Readers' Seniority

Tooltip

PhD / Post grad / Masters / Doc 16

64%

Professor / Associate Prof. 5

20%

Researcher 3

12%

Lecturer / Post doc 1

4%

Readers' Discipline

Tooltip

Medicine and Dentistry 10

45%

Pharmacology, Toxicology and Pharmaceut... 5

23%

Engineering 5

23%

Chemistry 2

9%

Article Metrics

Tooltip
Mentions
News Mentions: 3
Social Media
Shares, Likes & Comments: 70

Save time finding and organizing research with Mendeley

Sign up for free