Model-Based Recommendations for Optimal Surgical Placement of Epiretinal Implants

4Citations
Citations of this article
11Readers
Mendeley users who have this article in their library.
Get full text

Abstract

A major limitation of current electronic retinal implants is that in addition to stimulating the intended retinal ganglion cells, they also stimulate passing axon fibers, producing perceptual ‘streaks’ that limit the quality of the generated visual experience. Recent evidence suggests a dependence between the shape of the elicited visual percept and the retinal location of the stimulating electrode. However, this knowledge has yet to be incorporated into the surgical placement of retinal implants. Here we systematically explored the space of possible implant configurations to make recommendations for optimal intraocular positioning of the electrode array. Using a psychophysically validated computational model, we demonstrate that better implant placement has the potential to reduce the spatial extent of axonal activation in existing implant users by up to 55%. Importantly, the best implant location, as inferred from a population of simulated virtual patients, is both surgically feasible and is relatively stable across individuals. This study is a first step towards the use of computer simulations in patient-specific planning of retinal implant surgery.

Cite

CITATION STYLE

APA

Beyeler, M., Boynton, G. M., Fine, I., & Rokem, A. (2019). Model-Based Recommendations for Optimal Surgical Placement of Epiretinal Implants. In Lecture Notes in Computer Science (including subseries Lecture Notes in Artificial Intelligence and Lecture Notes in Bioinformatics) (Vol. 11768 LNCS, pp. 394–402). Springer Science and Business Media Deutschland GmbH. https://doi.org/10.1007/978-3-030-32254-0_44

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free